![Agilent Technologies E5070B User Manual Download Page 294](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065294.webp)
288
Chapter 7
Fixture Simulator
Determining Characteristics After Adding a Matching Circuit
Figure 7-5
Circuit models for defining matching circuits
NOTE
In the 2-port matching circuit embedding or the 2-port network de-embedding feature, if
the normalized impedance value specified in the user file (2-port Touchstone data file) is
different from the port reference impedance setting value of the analyzer, it is
automatically converted to adapt to the analyzer setting.
When a 2-port Touchstone data file is read in, data for up to 3202 frequency points are read
in using interpolation according to the measurement frequency point setting of the
analyzer. If the number of frequency points in the file is greater than 3202, excess data are
ignored.
For a network defined in the user file, it is assumed that port 1 is connected to the test port
and port 2 is connected to the DUT.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...