2-3
Installation
Preparation for Use
Preparation for Use
WA R N I N G
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD
REPLACE FUSE ONLY WITH SAME TYPE AND RATINGS (SEE
“Fuses” on page 2-3).
WA R N I N G
If this instrument is not used as specified, the protection
provided by the equipment could be impaired. This instrument
must be used in a normal condition only (in which all means for
protection are intact).
Power Requirements
The HP 37717C Communications Performance Analyzer requires a
power source of 90 V to 132 V ac and 198 V to 264 V ac at a frequency
between 47 Hz and 63 Hz (nominal).
Total power consumption is 450 VA (maximum).
The fuse rating for the power source is given in the following Table.
Fuses
Fuse Replacement
Only the ac line fuse located at the rear of the instrument may be
replaced by the operator.
WA R N I N G
ALL OTHER FUSE REPLACEMENT SHOULD ONLY BE
CARRIED OUT BY SUITABLY TRAINED SERVICE PERSONNEL
AWARE OF THE HAZARDS INVOLVED.
Line Voltage
Fuse Rating
HP Part Number
90V to 264V
5A Timed, 250V
2110-1120
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...