3-155
Performance Tests
Wander/Slips Measurement (Options UHN, [US9], A1M [A1Q], A1N
[A1R], A1P [A1S], A3L [A3M], A3V [A3W], A3N [A3P])
3. Press
and set up the
display as shown opposite.
4. Press
and set up the display
as shown opposite.
5. Connect the unbalanced 75
Ω
PDH
OUT port to the unbalanced 75
Ω
PDH IN
port and the unbalanced 75
Ω
TIMING
REF INPUT (Jitter Receiver Module)
simultaneously using the T-piece.
6. Press
and verify that the
display shows POSITIVE PEAK
WANDER and NEGATIVE PEAK
WANDER readings of 0.000
±
0.125 BITS.
7. Press
to stop the measurement.
8. Disconnect the PDH OUT port from the unbalanced 75
Ω
TIMING REF INPUT
and connect to the balanced 120
Ω
TIMING REF INPUT via the HP 15508C
Balanced to Unbalanced Converter.
9. Press
and select WANDER REFERENCE [120
Ω
BAL].
10. Press
and verify that the display shows POSITIVE PEAK
WANDER and NEGATIVE PEAK WANDER readings of 0.000
±
0.125 BITS.
11. Remove the input from the TIMING REF port and check that the display
indicates NO REF.
12. Press
to stop the measurement.
RECEIVE
RESULTS
RUN/STOP
RUN/STOP
RECEIVE
RUN/STOP
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...