
3-115
Performance Tests
STM-1/STM-4 Optical Interface (Options UKT, USN)
52. Reconnect the HP 37717C 1550 nm Optical Out Port to the HP 8153A Power
Meter via the HP 8157A Optical Attenuator.
53. Press
on the HP 37717C and select SIGNAL [STM-4 OPT]
[1550 nm].
54. Adjust the Optical Attenuator level to obtain a reading of -28dBm on the Power
Meter.
55. Recall the HP 37717C DEFAULT SETTINGS as shown on 3-2.
56. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
57. Disconnect the optical cable from the Power Meter and connect to the HP 37717C
Optical IN Port.
58. Press
on the HP 37717C and select SIGNAL [STM-4 OPT]
[1550 nm].
59. Press
then
to start the
measurement.
60. After 5 minutes, check that NO TROUBLE is displayed on the HP 37717C
display.
Press
to stop the measurement.
61. Repeat steps 50 through 60, selecting SIGNAL [STM-1 OPT] [1550 nm] in steps
53 and 58 and -34dBm in step 54.
62. Recall the HP 37717C DEFAULT SETTINGS as shown on Page 3-2.
63. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
64. Disconnect the optical cable from the HP 37717C STM1/4 Optical IN Port and
connect to the HP 83442A Optical/Electrical Converter input.
65. Connect the output from the Optical /Electrical Converter to the Oscilloscope
using the SMA/BNC adaptor and 50
Ω
BNC cable.
66. Set the Dual Power Supply to +12V and -12V, then connect to the HP 83442A
O/E Converter and switch on.
C AU T I O N
Take care to connect the supply correctly as incorrect voltage or polarity could result
in damage to the HP 83442A. Refer to HP 83442A Operating Instructions, HP p/n
5091-6448A.
TRANSMIT
SDH
TRANSMIT
SDH
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...