3-40
Performance Tests
PDH Frame Analysis (Options UKK, [USB])
14. Use the Frame Generator 8448 Kbit/s
and
keys to select and
enable AIS OUT.
15. Check that the HP 37717C Frame Loss and AIS alarm Led’s are ON.
16. Press the Frame Generator 8448 Kbit/s
key and check that the FRAME
(FAS) Error Count is 0 and all Alarm Led’s are now OFF.
17. Press
on the HP 37717C once to stop the measurement.
34 Mb/s Frame Analysis
1. Select SIGNAL [34 Mb/s] on the HP 37717C
display then return to
Display.
2. Set the HP 37729A Frame Generator parameters as follows:
OUTPUT: FREQ [34]; CODE [HDB3]
ALIGNMENT [34M]
SB [00]
FAS [1111010000]
Press the 34368 Kbit/s
key until all 34368 Kbit/s ERROR and AIS
Led’s are OFF.
3. Press
on the HP 37717C to start a measurement and check that all
HP 37717C Alarm Led’s are OFF. The Error Count and Error Ratio should be 0.
Error Detection
4. Press the 34368 Kbit/s
key until all 34368 Kbit/s ERROR and AIS
Led’s are OFF.
5. Use the Frame Generator 34368 Kbit/s
and
keys to select
and enable BIT ERROR.
6. Check that Errors and Remote alarm Led’s are ON.
7. Check that the HP 37717C counts Frame (FAS) Errors. The Code Error count
should be 0.
8. Press the Frame Generator 34368 Kbit/s
key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
Alarm Detection
9. Set the Frame Generator SB parameter to [10].
SELECT
ON/OFF
ON/OFF
RUN/STOP
RECEIVE
RESULTS
SELECT
RUN/STOP
SELECT
SELECT
ON/OFF
ON/OFF
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...