3-11
Performance Tests
PDH Transmitter Output (UKK, [USB], UKJ, [USA]
3. Press
and set up the
display as shown opposite.
4. Press
on the
oscilloscope.
5. Adjust the Oscilloscope Timebase and
Delay to position the positive peak pulse
amplitude at mid-pulse-width point in the
centre of the screen.
6. Verify the Pulse in Figure 3-1 meets the following criteria:
Pulse Amplitude
2.133V to 2.607V
Pulse Width
639 ns to 781 ns
Overshoot
≤
0.474V
Undershoot
≤
0.474V
Figure 3-1
704 kb/s Pulse Criteria
7. Adjust the Oscilloscope Delay to position the negative peak pulse amplitude at
mid-pulse-width point in the centre of the screen and verify that the negative pulse
meets the criteria listed in step 6.
704 kb/s Balanced Output
8. Connect the Balanced to Unbalanced Converter between the HP 37717C 120
Ω
Balanced SIGNAL OUT port and the Oscilloscope, leaving the 75
Ω
termination and
"T" connector in place.