3-171
Performance Tests
PDH Binary Interfaces (Option UH3, [US7])
5. Press
and set up the display as
shown opposite.
6. Press
and
set up the display as shown opposite.
7. Recall default settings on the Digital Transmission Analyzer and select: Interface
- NRZ BIN ECL Frequency - 170 MHz Pattern - 2
23
-1
8. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.
9. Select the following on the Digital Transmission Analyzer:
Frequency - 700 kHz
Pattern 2
15
-1
10. Select SIGNAL 2 Mb/s and PATTERN 2
15
-1 on the HP 37717C
and
displays.
11. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.
12. Select Interface - NRZ BIN TTL on the Digital Transmission Analyzer and
THRESHOLDS EXT CLOCK - TTL on the HP 37717C
BINARY
display.
13. Check that the HP 37717C Pattern Loss and Signal Loss indicators are not lit.