3-218
Performance Tests
ETSI/ANSI Internal Transmitter Clock Rates (Option 110)
ETSI/ANSI Internal Transmitter Clock Rates
(Option 110)
Specification
Description
A frequency Counter terminated in 75 ohms and connected to the Option 110 Trans-
mit Module front panel Clock Out Port is used to verify the frequency accuracy of
the transmitted data at all rates.
E3 (34.368Mb/s) Frequency Accuracy
1. Recall the HP 37717C Default Settings as shown on 3-2. Select TRANSMIT,
PDH/DSn, MAIN SETTINGS and set the SIGNAL to 34 Mb/s.
2. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 34367845.3 Hz and
34368154.7 Hz.
3. Set the FREQUENCY OFFSET to [USER OFFSET] [+20PPM].
4. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 34368532.7 Hz and
34368842.0 Hz.
5. Set the FREQUENCY OFFSET to [USER OFFSET] [-20PPM].
6. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 34367158.0 Hz and
34367467.3 Hz.
Rate
Frequency
Accuracy
Temperature
Stability
Ageing
Stability
DS1
1.544Mb/s
+/-0.7ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
DS3
44.736Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
E1
2.048Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
E3
34.368Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...