1-9
General Information
Operators Maintenance
2
Remove the adapters from the IN and OUT ports.
3
Using the blow brush with the brush removed blow through the ferrule of the
standard flexible connector and the adapter.
C AU T I O N
If the optical fibre of the fixed connector requires further cleaning this entails
disassembly of the module which should only be carried out by suitably trained
service personnel.
4
Apply some isopropyl alcohol to a piece of the cleaning paper and clean the barrel
of the adapter. Using a new piece of cleaning paper, clean the face of the adapter.
Repeat this operation, using a new piece of cleaning paper each time.
5
Lightly press the adhesive side of the tape provided against the front of the
adapter, then remove it quickly - repeat twice. This removes any particles of
cleaning paper which may be present.
6
Replace the adapters on the flexible connector.
Power Supply Fuse Replacement
See “Fuse Replacement” on page 2-3
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...