3-189
Performance Tests
ETSI/ANSI Transmitter Output Level and Waveshape (Option UKZ)
2. Select CHAN on the oscilloscope and set OFFSET to -390 mV.
3. Select TRIG on the oscilloscope and set to PATTERN H X X X.
4. Select TIMEBASE on the Oscilloscope and set DELAY to 300 nS. Adjust the
delay manually until the negative pulse is centred on the oscilloscope display.
5. If using the HP 54503A oscilloscope, use the following key sequence to select
and store the correct Telecom Mask.
a. WFORM SAVE key.
b. MASK DS1 new 1.5 Mb ANSI T1.102 softkey.
c. STORE to M1 + M2.
e. STORE INVERSE.
6. Compare the isolated negative pulse from the HP 37717C with the limits in the
selected Telecom Mask.
7. Ensure the waveform fits the mask.
N OT E
It may be necessary to visibly check the waveform fits the mask and manually adjust
amplitude and offset parameters on the oscilloscope to obtain a PASS on the Mask.
8. Measure the peak pulse amplitude at the mid pulse-width using the oscilloscope
and verify that this is between 1.984V and 2.976V.
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...