3-147
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
2. Select
and
setup the Test HP 37717C display as shown
opposite.
3. Connect up the equipment as shown in Figure 3-52.
Figure 3-52
Wander Generation
4. Set the Synthesizer to Frequency 2.048 MHz and Output Level 5V pk_pk.
TRANSMIT
JITTER