3-202
Performance Tests
Trigger Output (Option UKZ)
3. Select the HP 37717C TRANSMIT
page, select TEST FUNCTION and set
as follows;
4. Select the HP 37717C RECEIVE
page, select PHYSICAL LAYER and set
as follows;
5. Select the HP 37717C RECEIVE
page, select TRIGGERS and set as
follows;
6. Connect the Transmit Module (Option UKZ) 75 ohms Unbalanced Output to the
Receive Module (Option UKZ) 75 ohms Unbalanced Input.
7. Connect the Trigger Output port on the Receive Module (Option UKZ) to the
Oscilloscope Channel A. Terminate in 75 ohm.
8. Press AUTOSCALE on the Oscilloscope and ensure the displayed Trigger Output
pulse has TTL levels.