3-176
Performance Tests
ETSI/ANSI Internal Transmitter Clock Rates (Option UKZ)
ETSI/ANSI Internal Transmitter Clock Rates (Option
UKZ)
Specification
Description
A frequency Counter terminated in 75 ohms and connected to the Option UKZ Trans-
mit Module front panel Clock Out Port is used to verify the frequency accuracy of
the transmitted data at all rates.
DS3 Frequency Accuracy
1. Recall the HP 37717C Default Settings as shown on 3-2.
2. Connect the Clock Output port on the HP 37717C Option UKZ Transmit Module
to the Frequency Counter Channel A. Terminate in 75 ohm.
3. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 44735798.6 Hz and
44736201.4 Hz
4. Select the HP 37717C TRANSMIT page, select PHYSICAL LAYER and set
FREQUENCY OFFSET to [+20PPM].
5. Adjust the Frequency Counter ATTEN and Trigger Level to obtain a stable
reading and ensure that the frequency counter reading is between 44736693.4 Hz and
44737096.0 Hz.
6. Set the FREQUENCY OFFSET to [-20PPM].
Rate
Frequency
Accuracy
Temperature
Stability
Ageing
Stability
DS1
1.544Mb/s
+/-0.7ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
DS3
44.736Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
E1
2.048Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
E3
34.368Mb/s
+/-0.5ppm @25C
<=3ppm (0 to +50C)
<=1ppm/Year
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...