3-116
Performance Tests
STM-1/STM-4 Optical Interface (Options UKT, USN)
67. Press
on the HP 37717C and select SIGNAL [STM-4 OPT]
[1550 nm].
68. Press
on the oscilloscope and adjust the Timebase and Range to
obtain an STM-4 waveform.
69. Measure the amplitude of this STM-4 waveform using the oscilloscope and adjust
the Optical Attenuator until the amplitude is 150mV pk-pk.
N OT E
If the input power is greater than -8 dBm then the Optical/Electrical converter is liable
to saturate.
70. Disconnect the output of the O/E Converter from the oscilloscope and connect
instead to the HP 37717C STM1/4 Optical Module Monitor Input.
71. Press
on the HP 37717C
and set up the display as shown opposite.
72. Press
then
to start the measurement.
73. After 5 minutes, check that NO
TROUBLE is displayed on the HP 37717C
display. Press
to
stop the measurement.
74. Recall the HP 37717C DEFAULT SETTINGS as shown on 3-2.
75. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
76. Repeat steps 62 through 75, selecting SIGNAL [STM-1 OPT] [1550 nm] in step
67 and [STM-1 MON] in step 71.
77. Disconnect all the test equipment.
TRANSMIT
SDH
AUTOSCALE
RECEIVE
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...