1-47
General Information
Specifications
The HP 37717C must be connected
back to back in order to perform a
calibration cycle before making a Jitter
Transfer measurement.
The HP 37717C must have been
switched on for 1 hour before starting a
calibration cycle.
The climatic conditions must remain
stable from switch-on to end of
measurement.
It is recommended that the maximum
test period (start of calibration to end of
measurement) does not exceed 90
minutes.
Test Period = Delay Time + Dwell
Time + 5 Seconds X Number of Points
X 2 (Calib Measurement).
Peak Jitter Measurement
These ranges cover the measurements
range specified in O.171, Table3/O.171.
Note that the receiver will be able to
handle transients of up to 24UI(peak)
without losing lock, although the
measurement will be out of range.
On Losing Lock the Jitter Loss LED is
illuminated. If Lock is maintained but
the measurement is out of range then a
Status Message “Jitter Out of Range” is
displayed on the Status Line.
*
7Hz For Rev.A of Options UHK, A1M, A1N, A1P
**
Meets ITU O.171 Table 4.
Receiver Ranges
Range
Lower
UI pp
Upper
UI pp
1.6
0
1.6
16
0
16
Peak Measurement Bandwidth
Bit Rate (Kb/s)
Bandwidth
Min.
*
Max
Range 1.6UI
**
Range 16UI
2048
2Hz
100 kHz
50 kHz
8448
2Hz
400 kHz
100 kHz
34368
2Hz
800 kHz
400 kHz
139264
2Hz
3.5 MHz
800 kHz
155520
2Hz
1.3 MHz
800 kHz
622080
2Hz
5.0 MHz
800 kHz
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...