1-50
General Information
Specifications
Internal Filters:
A number of internal filters are
available to band limit the demodulated
jitter signal before jitter amplitude
measurement. These filters are as
specified in ITU O.171.
*
Filter tested to be within 10% of nominal.
**
Minimum frequency for the 3dB points.
***
Options A3L, A3V and A3N only.
For high pass filters the slope below the
3 dB point is 20 dB per decade. For the
low pass filters the slope above the 3 dB
point is 60 dB per decade.
The filters can be selected as follows
• OFF
• LP
• HP1
• HP2
• LP HP1
• LP + HP2
• 12 kHz
The 12 kHz filter is not characterised
for pk-pk measurements, hence
operation is not specified. (This filter is
not recommended in any relevant
ITU-T pk-pk measurement standard).
RMS Jitter Measurement (Options
A3L, A3V and A3N only):
This measurement is available at all
interface rates.
Filters: 12 kHz HP filter selectable.
If HP1, HP2 and/or LP are enabled,
then they will affect the jitter signal
measured by the RMS receiver.
RMS Results Resolution: The RMS
ranges are linked to the selection for
Peak. No separate selection will exist.
Peak Range 1.6 UI: - RMS range 0.5,
steps of 0.002 UI.
Peak Range 16 UI:- RMS Range 5,
steps of 0.02 UI.
Nominal 3 dB Corner Frequencies
Filter (kb/s)
HP1
*
(high pass) (Hz)
HP2
*
(high pass) (kHz)
LP
**
(low pass) (kHz)
2048
20
18
100
8448
20
80
400
34368
100
10
800
139264
200
10
3500
155520
500
65
***
1300
622080
1000
250
***
5000
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...