1-27
General Information
Specifications
† APS threshold N errors in T ms where
0
≤
N
≤
1920 (STM-3) and 10 ms
≤
T
≤
10000 s, in decade steps.
* For SONET stand-alone operation,
bulk-filled payloads and DS1, DS3
mapped payloads only.
For bit error rates supported with other
payloads refer to the DSn/PDH test
option for details.
Alarm generation: LOS, LOF, SEF,
AIS-L, RDI-L.
STS-3c SPE AIS-P, STS-3c SPE RDI-
P, STS-3c SPE LOP, STS-3c SPE path
unequipped.
STS-1 SPE AIS-P, STS-1 SPE RDI-P,
STS-1 SPE LOP, STS-1 SPE path
unequipped.
VT6 path AIS, VT6 RDI-V, VT6 LOP,
VT6 path unequipped, VT6 H4 LOM
(loss of multiframe).
VT2 path AIS, VT2 RDI-V, VT2 LOP,
VT2 path unequipped,
Error type
Single
Rate 10
-N
Comments
Frame A1A2
*
N in 4 frame words
CV-S (B1)
*
4 - 9
CV-2 (B2)
*
3 - 9
REI-L
*
3 - 9
STS-3c SPE CV-P (B3)
*
4 - 9
STS-3c SPE REI-P
*
4 - 9
STS-3c SPE IEC-P
*
4 - 9
STS-1 SPE CV-P (B3)
*
4 - 9
STS-1 SPE REI-P
*
4 - 9
STS-1 SPE IEC-P
*
4 - 9
VT6 CV-P (V5)
*
4 - 9
VT6 REI-P
*
5 - 9
VT2 CV-P (V5)
*
3 - 9
VT2 REI-P
*
4 - 9
VT1.5 CV-P
*
3 - 9
VT1.5 REI-P
*
4 - 9
Bit error *
*
3 - 9
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...