3-41
Performance Tests
PDH Frame Analysis (Options UKK, [USB])
10. Check that the HP 37717C Remote Alarm Led is ON. Error Count and Error
Ratio should be 0.
11. Reset the Frame Generator SB parameter to [00].
FAS Error and AIS
12. Use the Frame Generator 34368 Kbit/s
and
keys to select
and enable FAS ERROR.
13. Check that the HP 37717C Frame Loss alarm Led is ON.
14. Press the Frame Generator 34368 Kbit/s
key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
15. Use the Frame Generator 34368 Kbit/s
and
keys to select
and enable AIS OUT.
16. Check that the HP 37717C Frame Loss and AIS alarm Led’s are ON.
17. Press the Frame Generator 34368 Kbit/s
key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
18. Press
on the HP 37717C once to stop the measurement.
140 Mb/s Frame Analysis
1. Select SIGNAL: [140 Mb/s] on the HP 37717C
display then return to
Display.
2. Set the HP 37729A Frame Generator parameters as follows:
OUTPUT: FREQ [139]; CODE [CMI]
ALIGNMENT [140M]
SB [0000]
FAS [111110100000]
Press the 139264 Kbit/s
key until all the 139264 Kbit/s ERROR and
AIS Led’s are OFF.
3. Connect the Frame Generator CMI/NRZ output to the HP 37717C 75
Ω
SIGNAL
IN port.
4. Press
on the HP 37717C to start a measurement and ensure all
Alarm Led’s are OFF. Check that the HP 37717C Error Count and Error Ratio is 0.
SELECT
ON/OFF
ON/OFF
SELECT
ON/OFF
ON/OFF
RUN/STOP
RECEIVE
RESULTS
SELECT
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...