2-16
Installation
Connecting Accessories
Hewlett-Packard Interface Bus
The HP 37717C Communications Performance Analyzer (Option A3B or A3D) is
connected to the HP-IB by means of an appropriate HP-IB cable.The HP-IB
interconnecting cables available are listed in the following table.
HP-IB Interconnecting Cables
To achieve interface design performance standards, restrictions are placed on the
HP-IB system cable lengths. These restrictions allow the bus interface electronics to
maintain correct line voltage levels and timing relationships.
When connecting an HP-IB system the following rules should be observed:
The total HP-IB cable length used must be less than or equal to 20 meters (65.6
feet).
The total HP-IB cable length used must be less than or equal to 2 meters (6 feet)
×
the total number of devices connected to the bus.
A standard HP-IB connector is provided on the instrument . The connections are
shown in ther following figure. The mating connector part number is HP 1251-0293
or Amphenol 57-30240.
Length
Accessory Number
1 meter
HP 10833A
2 meters
HP 10833B
4 meters
HP 10833C
0.5 meter
HP 10833D
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...