3-42
Performance Tests
PDH Frame Analysis (Options UKK, [USB])
Error Detection
5. Use the Frame Generator 139264 Kbit/s
and
keys to select
and enable BIT ERROR.
6. Check that Errors and Remote alarm Led’s are ON.
7. Check that the HP 37717C counts Frame (FAS) Errors. The Code Error count
should be 0.
8. Press the Frame Generator
key and check that the FRAME (FAS) and
CODE error count is 0 and the Alarm Led’s are OFF.
Alarm Detection
9. Set the Frame Generator SB parameter to [1000].
10. Check that the HP 37717C Remote Alarm Led is ON. Check that the Error Count
and Error Ratio is 0.
11. Reset the Frame Generator SB parameter to [0000].
FAS Error and AIS
12. Use the Frame Generator 139264 Kbit/s
and
keys to select
and enable FAS ERROR.
13. Check that the HP 37717C Frame Loss alarm Led is ON.
14. Press the Frame Generator 139264 Kbit/s
key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
15. Use the Frame Generator 139264 Kbit/s
and
keys to select
and enable AIS OUT.
16. Check that the HP 37717C Frame Loss and AIS alarm Led’s are ON.
17. Press the Frame Generator 139264 Kbit/s
key and check that the
FRAME (FAS) Error Count is 0 and all Alarm Led’s are now OFF.
18. Press
on the HP 37717C once to stop the measurement.
19. Disconnect the test equipment.
SELECT
ON/OFF
ON/OFF
SELECT
ON/OFF
ON/OFF
SELECT
ON/OFF
ON/OFF
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...