3-145
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
External Jitter Generation
1. Select
and setup the HP
37717C display as shown opposite.
2. Select
and
setup the display as shown opposite.
3. Set the Spectrum Analyzer as follows:
Centre Frequency - 2048 kHz
Frequency Span - 25 kHz
Reference Level - 0 dBm
Sweep Time - 1.0 s
Video Bandwidth - 1 kHz
Resolution Bandwidth - 100Hz
4. Set the Synthesizer to Frequency 10 kHz and minimum Output Level.
TRANSMIT
TRANSMIT
JITTER
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...