3-83
Performance Tests
External MTS Clock (Options US1, [US5], A1T, [A1U], A3R [A3S], 120)
5. Press
and set up the display
as shown opposite.
6. Press
;
.
7. Press
, check that the
display reads NO TROUBLE and all the
Alarm leds are off.
8. Press
to halt the
measurement.
9. Disconnect the HP 3335A and check
that the CLOCK LOSS led is on, but the
display reads NO TROUBLE.
External MTS - 2MHz Data
10. Select CLOCK : EXTERNAL [MTS 2M DATA].
11. Set the HP 3335A to output a 1.024MHz, 2V p-p signal.
12. Connect the HP 3335A to the 75
Ω
2M REF IN port.
13. Press
to start the measurement and check that the RESULTS display
reads NO TROUBLE and all the Alarm leds are off.
14. Disconnect the HP 3335A and check that the CLOCK LOSS led is on.
15. Connect the HP 3335A via the 75
Ω
/120
Ω
Converter to the 120
Ω
2M REF IN
port.
16. Press
and check that the RESULTS display reads NO TROUBLE
and all the Alarm leds are off.
BITS 1.5 Mb/s (Option 120 only)
17. Select CLOCK : EXTERNAL [BITS : 1.5M].
18. Set the HP 3335A to output a 772kHz, 2V p-p signal.
19. Connect the HP 3335A to the 100
Ω
1.5M REF IN port.
20. Press
and check that the RESULTS display reads NO TROUBLE
and all the Alarm leds are off.
RECEIVE
RESULTS
TROUBLE SCAN
RUN/STOP
RUN/STOP
RUN/STOP
RUN/STOP
RUN/STOP
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...