3-107
Performance Tests
STM-1/STM-4, 1550nm Optical Interface (Option URU)
10. Press
on the HP 37717C
and set up the display as shown opposite.
11. Press
on the HP 37717C
and select SIGNAL [STM-4 OPT].
12. Press
then
to start the measurement.
13. After 5 minutes, check that NO
TROUBLE is displayed on the HP 37717C
display.Press
to
stop the measurement.
14. Press
on the HP 37717C and select SIGNAL [STM-1].
15. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
16. Reconnect the HP 37717C STM1/4 Optical Out Port to the HP 8153A Power
Meter via the HP 8157A Optical Attenuator.
17. Press
on the HP 37717C and select SIGNAL [STM-4 OPT].
18. Adjust the Optical Attenuator level to obtain a reading of -26dBm on the Power
Meter.
19. Press
on the HP 37717C and select SIGNAL [STM-1].
20. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
21. Disconnect the optical cable from the Power Meter and connect to the HP 37717C
STM1/4 Optical Rx Input.
22. Press
on the HP 37717C and select SIGNAL [STM-4 OPT].
23. Press
then
to start the
measurement.
24. After 5 minutes, check that NO TROUBLE is displayed on the HP 37717C
display. Press
to stop the measurement.
25. Press
on the HP 37717C and select SIGNAL [STM-1].
26. Verify that the Optical Module LASER ON led is NOT lit indicating that the laser
is disabled before continuing.
27. Disconnect the optical cable from the HP 37717C STM1/4 Optical Rx Input and
connect to the HP 83442A Optical/Electrical Converter input.
RECEIVE
TRANSMIT
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
TRANSMIT
SDH
TRANSMIT
SDH
TRANSMIT
SDH
TRANSMIT
SDH
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
TRANSMIT
SDH
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...