3-127
Performance Tests
Transmitted Jitter Amplitude Accuracy (Option UHK)
3. Make the following key sequence on
the HP 37717C to obtain the special
display.
Press
;
;
;
;
;
;
;
. .
Press
until
appears in the softkey menu.
Press
and select
MODULE PDH Module.
Set the VCO Control Mode to FIXED.
C AU T I O N
The sequence above must be performed each time the power is cycled on the HP
37717C, as all
parameters adopt DEFAULT values when power
is cycled.
Procedure
Transmitted Jitter Amplitude Accuracy on Range 1
1. Setup the HP 37717C
display as shown opposite.
MODULE DEBUG
OTHER
MORE
MORE
MORE
OTHER
MORE
MODULE DEBUG
MODULE DEBUG
MODULE DEBUG
TRANSMIT