3-53
Performance Tests
SPDH Receiver Monitor Levels (Options UKJ, [USA], UKL, [USC], UKN,
[USE])
15. Ensure that NO TROUBLE is displayed on the
display after 30
seconds. Press
to stop the measurement.
16. Remove Cable Simulator #1 from the signal path.
17. Press
and select EQUALIZATION [OFF].
18. Press
;
. Press
to start the
measurement.
19. Ensure that NO TROUBLE is displayed on the
display after 30
seconds. Press
to stop the measurement.
20. Connect the second 3dB Fixed Attenuator in the signal path to give a total path
attenuation of 29.8dB (7.6+3+3+6+6+4.2).
21. Press
and set MONITOR LEVEL [30dB].
22. Press
. Press
to start the
measurement.
23. Ensure that NO TROUBLE is displayed on the
display after 30
seconds. Press
to stop the measurement.
24. Press
and select EQUALIZATION [ON].
25. Connect Cable Simulator #1 between the PDH OUT Port and the Matching Pad.
26. Press
. Press
to start the
measurement.
27. Ensure that NO TROUBLE is displayed on the
display after 30
seconds. Press
to stop the measurement.
8.448 Mb/s
28. Repeat steps 1 to 26 (20dB, 26dB and 30dB tests) with the HP 37717C
and
displays set to SIGNAL [8 Mb/s] and Cable Simulator
#2 fitted in place of Cable Simulator #1.
34.368 Mb/s
29. Repeat steps 1 to 18 (20dB and 26dB tests) with the HP 37717C
and
displays set to SIGNAL [34 Mb/s] and Cable Simulator #3 fitted in place
of Cable Simulator #2.
RESULTS
RUN/STOP
RECEIVE
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
RECEIVE
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
RECEIVE
RESULTS
TROUBLE SCAN
RUN/STOP
RESULTS
RUN/STOP
TRANSMIT
RECEIVE
TRANSMIT
RECEIVE
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...