3-237
Performance Tests
Performance Test Record
Performance Test Record
Hewlett-Packard Model 37717C Communications Performance Analyzer
Location:
Serial No.:
Tested by:
Temperature:
Certified by:
Humidity:
Date:
Table 11
Performance Test Record
Page
No.
Test Description
Result
Min Actual Max
Self Test
3-3
Step 3
All Tests
Pass/Fail
PDH Internal Transmitter Clocks
3-7
Step 5:
704 kb/s (UKK only)
351.99753 kHz
352.00246 kHz
Step 6:
2 Mb/s
1.0239928 MHz
1.0240072 MHz
Step 7:
8 Mb/s
4.2239704 MHz
4.2240296 MHz
Step 8:
34 Mb/s
17.1838797 MHz
17.1841203 MHz
Step 9:
140 Mb/s
69.6315126 MHz
69.6324874 MHz
Frequency Offset Bit Rates
140 Mb/s
3-8
Step 11:
+ 15 ppm
69.63255705 MHz
69.63353191 MHz
Step 12:
- 15 ppm
69.63064218
MHz
69.63126886
MHz
Step 13:
+ 100 ppm
69.63864986
MHz
69.63927654
MHz
Step 14:
- 100 ppm
69.62472346 MHz
69.625535014 MHz
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...