3-80
Performance Tests
SDH Frequency Offsets (Options US1, [US5], A1T, [A1U], A3R [A3S],
120)
3. Press
and set up the
display as shown opposite.
4. Make the following key sequence on
the HP 37717C to obtain the special
display.
Press
;
;
;
;
;
;
;
. .
Press
until
appears in the softkey menu.
Press
and set up the
display as shown opposite.
C AU T I O N
When using the
display, ONLY modify the STM-1 TEST
PATTERN. Altering other parameters can damage instrument firmware - exit this
display after setup to eliminate any possibility of accidental modification.
5. Adjust the Frequency Counter Trigger Level to obtain a stable reading.
6. Take note of the measured frequency (MF).
7. Select the frequency offset settings given in Table 3-4 and verify the frequency at
each step against the measured frequency (MF) noted in step 6.
TRANSMIT
MODULE DEBUG
OTHER
MORE
MORE
MORE
OTHER
MORE
MODULE DEBUG
MODULE DEBUG
MODULE DEBUG