3-68
Performance Tests
PDH Receiver Monitor Levels (Special Option 808)
PDH Receiver Monitor Levels (Special Option 808)
Specification
Description
Option 808 provides an equalized Monitor input. This is checked by attenuating the
synthesizer output level by an amount equivalent to the extra gain provided by the
Monitor Amplifier. The synthesizer is set to generate a sinewave at half the bit rate
which corresponds to a ternary all ones signal. The receiver should sync up with no
errors if PATTERN [ALL ONES] is selected.
Equipment Required
Procedure
1. Recall the HP 37717C DEFAULT SETTINGS as shown on 3-2.
2. Connect up the equipment as shown in Figure 3-32.
3. Set the oscilloscope to: TERMINATION - 1 M
Ω
PROBE - 1:1
Bit Rate
Nominal Loss
Equalization
at 1/2 Bit Rate
704 kb/s
20dB Flat
3dB
2.048 Mb/s
20dB Flat
3dB
8.448 Mb/s
20dB Flat
3dB
34.368 Mb/s
20dB Flat
6dB
139.264 Mb/s
20dB Flat
6dB
Synthesizer
: HP 3335A option 001 (75
Ω
)
Oscilloscope
: HP 54503A
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...