3-24
Performance Tests
Multiple PDH Transmitter Outputs (Options UHC, [US6])
6. Verify the Pulse meets the following criteria:
Pulse Amplitude
2.133V to 2.607V
Pulse Width
639 ns to 781 ns
Overshoot
≤
0.474V
Undershoot
≤
0.474V
Figure 3-15
704 kb/s Pulse Criteria
7. Repeat steps 5 and 6 with the Oscilloscope Channel 1 connected to SIGNAL
OUT 3 and SIGNAL OUT 4 in turn.
8. Adjust the Oscilloscope Timebase and Delay to position the negative peak pulse
amplitude at mid-pulse-width point in the centre of the screen and verify that the
negative pulse meets the criteria listed in step 6.
9. Repeat step 8 with the Oscilloscope Channel 1 connected to SIGNAL OUT 2 and
SIGNAL OUT 3 in turn.
704 kb/s Additional Outputs Delay
10. Connect the PDH SIGNAL OUT 1 port to the Oscilloscope Channel 1. Set the
Oscilloscope termination to 1M
Ω
.
11. Connect the PDH SIGNAL OUT 2 port to the Oscilloscope Channel 2. Set the
Oscilloscope termination to 1M
Ω
.
12. Check that the pulse on Channel 2 is 4 bits delayed with respect to the pulse on
Channel 1.
13. Disconnect the SIGNAL OUT 2 port from the oscilloscope and connect the
SIGNAL OUT 3 port to the oscilloscope Channel 2.
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...