3-179
Performance Tests
ETSI/ANSI Transmitter Output Level and Waveshape (Option UKZ)
ETSI/ANSI Transmitter Output Level and Waveshape
(Option UKZ)
Specifications
Description
This test ensures the transmitter output level and pulse shape meet the required spec-
ifications at E1,E3,DS1 and DS3 rates. The Transmitter output is connected to an os-
cilloscope and the waveshape compared with the predefined masks stored in the
oscilloscope memory. The signal levels are also measured using the oscilloscope.
Rate
Frequency
Level
Waveshape
DSX-1
1.544Mb/s
3.0V +/-20%
Fits mask T1.102-1993
DS1-LO
1.544Mb/s
As DSX-1 with 655'
ABAM Cable
DS3-HI
44.736Mb/s
0.9V pk (nominal)
DSX-3
44.736Mb/s
560mV pk (nomi-
nal)
Fits mask T1.102-1993
DS3-900
44.736Mb/s
330mV pk (nomi-
nal)
E1 (bal)
2.048Mb/s
3.00V +/-10%
As per ITU rec G703
E1 (unbal)
2.048Mb/s
2.37V +/-10%
As per ITU rec G703
E3
34.368Mb/s
1.0V +/-10%
As per ITU rec G703