B-2
Appendix B - Fitting/Calibrating/Testing New Modules
* 37717-60020/21 may be used (Structured PDH Modules). ** 37714-60012/
60013 may be used (STM1/4 Modules).
C AU T I O N
If 37714-60012 (Option URU) is fitted, the optical output must be attenuated by 10
dB before connecting to the Optical input of any other 37717C optical module.
Failure to do this could result in damage to the HP 37717C. Table 3-1 lists a suitable
attenuator.
STM-1 Electrical Jitter
Receiver
A1M,
[A1Q] A3L
[A3M]
Jitter Tx PDH
Tx/Rx SDH Tx/
Rx
37717-60010
37714-60044*
37714-60048
Jitter Tx
STM-1 Optical Jitter Rx
A1N, [A1R]
A3V [A3W]
Jitter Tx PDH
Tx/Rx SDH Tx/
RxS TM-1 Optic
Tx
37717-60010
37714-60044*
37714-60048
37714-60011**
Jitter Tx
STM-1/4 Optical Jitter
Receiver
A1P, [A1S]
A3N [A3P
Jitter Tx PDH
Tx/Rx SDH Tx/
Rx STM1/4
Optic Tx
37717-60010
37714-60044*
37714-60048
37714-60013**
Jitter Tx
Additional PDH Outputs
UHC, [US6]
PDH Tx/Rx
37714-60044*
No
Table B-1
Additional Modules Required for Performance Tests
Module Under Test
Option
Under Test
Module(s)
Required
Part Number
Calibration
Required?
Summary of Contents for 37717C
Page 2: ...HP 37717C Communications Performance Analyzer Calibration Manual ...
Page 5: ...2 ...
Page 21: ...xviii Contents ...
Page 109: ...2 20 Installation Additional Precautions for Service Engineers ...
Page 312: ...3 203 Performance Tests Trigger Output Option UKZ Figure 3 57 Monitor Input Test Setup ...
Page 375: ...3 266 Performance Tests Performance Test Record ...
Page 379: ...4 4 Installation Current Previous Terminology ...
Page 392: ...B 9 Appendix B Fitting Calibrating Testing New Modules Typical Jitter Transmit Spectrum ...
Page 393: ...B 10 Appendix B Fitting Calibrating Testing New Modules ...