C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-51
set temp std
value
This command sets the standard temperature to
value
, where
value
is a
floating-point number representation of standard temperature between 0 to
25 °C. The following example sets the standard temperature to 15 °C.
Send:
set
temp
std
Receive:
set temp std 15 ok
temp th
This command reports the heater temperature threshold for the sample.
The following example reports that the temperature threshold is 30 °C.
Send:
temp
th
Receive:
temp th 30 degC
set temp th
value
This command sets the heater temperature threshold for the sample to
value
, where
value
is a floating-point number representation of temperature
threshold between 0 to 70 °C. The following example sets the temperature
threshold to 35 °C.
Send:
set temp th 35
Receive:
set temp th 35 ok
contrast
This command reports the screen’s level of contrast. The following example
shows the screen contrast is 55%, according to
Table B–6
.
Send:
contrast
Receive:
contrast 11: 55%
set contrast
level
This command sets the screen’s
level
of contrast, according to
Table B–6
.
The following example sets the contrast level to 50%.
Send:
set
contrast
10
Receive:
set contrast 10 ok
Table B–6.
Contrast Levels
Level
Contrast Level
0 0%
1 5%
2 10%
3 15%
4 20%
Hardware
Configuration
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......