Operation
Calibration Factors Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-17
NEPHELOMETER BKG:
>NEPH MODE
VALUES
RESTORE PREV VALUES
RANGE AVG DIAGS ALARM
The Nephelometer Mode screen allows the user to place the nephelometer
into one of three modes. Available modes are Sample, Zero, and Stdby.
The sample mode begins photometric detection. The zero mode begins a
zeroing process that will begin with a filter tape change and the standby
mode (Stdby) mode will stall any measurement by the nephelometer.
●
In the Main Menu, choose Calibration Factors > Neph Bkg >
Neph
Mode
.
NEPHELOMETER MODE:
CURRENTLY: STDBY
SET TO: SAMPLE ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Values screen displays the optional background for each range of the
nephelometer. These values can be control-charted by the user to help
evaluate any optional drift.
●
In the Main Menu, choose Calibration Factors > Neph Bkg >
Values
.
BEFORE AFTER
1 0.00278 0.00277
2 0.00278 0.00277
3 0.01119 0.01122
4 0.04452 0.04459
5 0.17809 0.17808
RANGE AVG DIAGS ALARM
The Restore Previous Values screen is used to restore the prior background
values should a poor zero occur.
●
In the Main Menu, choose Calibration Factors > Neph Bkg >
Restore Prev Values
.
Nephelometer Mode
Values
Restore Previous Values
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......