Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-73
RELATIVE HUMIDITY:
ACTUAL MIN: 5.0 %
SET MIN TO: 5.5 %?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Sample RH screen displays the current sample relative humidity value
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 5 to 95%. If the sample RH reading goes beyond the minimum
or maximum limit, an alarm is activated and the alarm (bell) icon appears
in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > RH/Termperature Alarms >
Sample RH
.
SAMPLE RH:
ACTUAL 0.0 %
>MIN 5.0 %
MAX 95.0 %
RANGE AVG DIAGS ALARM
The Minimum Sample RH alarm limit screen is used to change the
minimum ambient RH alarm limit. The minimum and maximum sample
RH screens function the same way.
●
In the Main Menu, choose Alarms > RH/Temperature Alarms >
Sample RH >
Min
or
Max
.
SAMPLE RH:
ACTUAL MIN: 5.0 %
SET MIN TO: 35 %?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Ambient Temperature screen displays the current ambient temperature
value and sets the minimum and maximum alarm limits. Acceptable alarm
limits range from -30 °C to 60 °C. If the ambient temperature reading goes
beyond the minimum or maximum limit, an alarm is activated and the
Sample RH
Min and Max Sample RH
Limits
Ambient Temperature
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......