Calibration
Calibration Procedure
4-14
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Service > Neph Calibration >
Neph Temp
.
CALIBRATE NEPH TEMP:
NEPH TEMP: 0.0
o
C
OFFSET: 0.0
o
C
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Source Level screen allows the user to view and
calibrate the source level. This should be within 60-70 mA. If it is outside
of this range, please adjust the % SRC Level until the IRED reads 65 mA
●
In the Main Menu, choose Service > Neph Calibration >
Neph Src
Level
.
SOURCE LEVEL SET:
IRED: 65 mA
SRC LEVEL: 32 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Nephelometer Source
Level