Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-99
SOURCE LEVEL SET:
IRED: 70 mA
SRC LEVEL: 32 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Analog Output Calibration menu provides access to the 6 voltage
channels and 6 current channels for calibration. Current channels are
visible only if the I/O expansion board is installed. The analog output
calibration menu is visible only when the instrument is in service mode.
For more information on the service mode, see “
Service Mode
” earlier in
this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Analog Out Calibration
.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a volt
meter to the output and adjust the output until it reads 0.0 V on the meter
for a voltage channel, or either 0 or 4 mA for a current channel (depending
on the selected range). See the “set output to:” field on the display.
●
In the Main Menu, choose Service > Analog Out Cal > select
Channel >
Calibrate Zero
.
Analog Output Calibration
Analog Output Calibrate
Zero
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......