C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-41
Send:
set thermal mass coef 20.5
Receive:
set thermal mass coef 20.5 ok
vac mass coef
This command reports the current vacuum mass coefficient. The following
example reports that the vacuum mass coefficient is 2.0.
Send:
vac
mass
coef
Receive:
vac mass coef 2.000000
set vac mass coef
value
This command sets the vacuum mass coefficient to user-defined values to
value
, where
value
is a floating-point representation of the coefficient. The
following example sets the vacuum mass coefficient to 3.1.
Send:
set vac mass coef 3.1
Receive:
set vac mass coef 3.1 ok
neph cal bkg
This command sets the nephelometer into an auto-zero calibration routine.
The following example set this command.
Send:
set neph cal bkg
Receive:
set neph cal bkg ok
neph coef
This command reports the nephelometer span coefficient set at the factory.
The following example reports that the nephelometer span coefficient is
388.000.
Send:
neph
coef
Receive: neph
coef
388.000
set neph coef
value
This command sets the nephelometer span coefficient to user-defined
values to
value
, where
value
is a floating-point representation of the
coefficient. The following example sets the nephelometer span coefficient
to 250.00
Send:
set neph coef 250
Receive:
set neph coef 250 ok
neph reset bkg
This command will reset all measured background values to zero. The
following example sets this command.
Send:
set neph reset bkg
Receive:
set neph reset bkg ok
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......