Operation
Range Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-11
CONC UNITS:
CURRENTLY: mg/m3
SET TO:
3
μ
g/m3 ?
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The SHARP Range screen defines the concentration range of the analog
outputs. For example, a SHARP range of 0–100 μg/m
3
sets the analog
output to concentrations between 0 and 100 μg/m
3
by default.
The display shows the current SHARP range. The next line of the display is
used to change the range.
Table 3–2
lists the standard ranges.
●
In the Main Menu, choose Range >
SHARP Range
.
SHARP RANGE:
CURRENTLY: 100.0
SET TO: 1000 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3–2.
Standard Ranges
mg/m
3
μ
g/m
3
0.1 100
1.0 1000
2.0 2000
3.0 3000
5.0 5000
10.00 10000
C1 C1
C2 C2
C3 C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “Set Custom Ranges” below.
SHARP Range
Содержание 5030i SHARP