C-Link Protocol Commands
Diagnostics
B-24
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
baro pres
This command reports the current barometric pressure, in mmHg. The
following example reports that the barometric pressure is 722 mmHg.
Send:
baro
pres
Receive:
baro pres 722.0 mmHg
board temp
This command reports the current board temperature, in degrees C. The
following example reports that the board temperature is 15 °C.
Send:
board
temp
Receive:
board temp 15.0 degC
crn
This command reports the current radon concentration for the detector.
The following example reports that the radon concentration is 1.2 Bq/m
3
.
Send:
crn
Receive: crn
1.20
Bq/m3
det status
This command reports the current mass for the detector. The sequence is:
,
,
C
,
REF
. Each value is separated by a space.
Send:
det
status
Receive:
det status 1 9657 9600 4623
diag volt det
This command reports the diagnostic voltage measurements on the
detector board. The voltage is positive 5.
Send:
diag
volt
det
Receive:
diag volt det 4.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
iob
Receive:
diag volt iob 24.10 4.90 3.20 -3.20
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......