Operation
Instrument Controls Menu
3-22
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
START DATE AND TIME:
01 NOV 2008 00:00 ?
SETTING: DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
MANUAL:
>BENCH(S) CLOSE
TAPE STOP
TOGGLE VALUE
RANGE AVG DIAGS ALARM
The Mass Limit screen allows the user to set the mass limit. The default is
1,500 μg (or 1.5 mg) and is specified as part of any U.S. EPA and TUV
approvals. However, the limit can be increased as much as 5,000 μg for
non-regulatory monitoring.
●
In the Main Menu, choose Instrument Controls > Filter Tape
Control >
Mass Limit
.
MASS LIMIT:
CURRENTLY: 1500
SET TO: 150
1
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Next Time screen is used to view and set the initial date and time (24-
hour format) for the next filter tape change.
●
In the Main Menu, choose Instrument Controls > Filter Tape
Control >
Next Time
.
●
●
●
●
●
The Period screen allows the user to set the period. The next filter tape
change would be scheduled after the number of hours the period is set for.
The official tape change for TUV and U.S. EPA for PM
2.5
and PM
10
is set
for 8 hours.
Mass Limit
Next Time
Period
START DATE AND TIME:
05 MAY 2008 00:00
PRESS
TO EDIT
RANGE AVG DIAGS ALARM
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......