Servicing
Pressure Board Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-19
3.
Disconnect the pressure board from the divider panel.
4.
Remove the four pressure board assembly retaining screws and remove
the pressure board assembly (
Figure 7–9
).
Figure 7–9.
Replacing the Pressure Board Assembly
5.
To install the pressure board assembly, follow the previous steps in
reverse.
Unhook Pressure
Cable from
Interface Board
“Press” Board
Remove Screws (4)
Remove Plumbing