Operation
Diagnostics Menu
3-62
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
FLOW: 16.67 LPM
RANGE AVG DIAGS ALARM
The Detector Status screen (read only) displays the current sample mass
reading, alpha counts, and beta counts.
Note
A mass value will only be calculated when the flowrate is within
acceptable limits. However, when in Service Mode, a mass value will be
calculated if the pump is turned OFF.
▲
●
In the Main Menu, choose Diagnostics >
Detector Status
.
DETECTOR STATUS:
MASS 0 ug
ALPHA 2 1/sec
BETA 13577 1/sec
BETA CORR 13577 1/sec
BETA ZERO 6500 1/sec
RANGE AVG DIAGS ALARM
The Nephelometer Status screen (read only) displays the current
nephelometer reading.
●
In the Main Menu, choose Diagnostics >
Nephelometer Status
.
NEPHELOMETER STATUS:
NEPH TEMP 31.0
o
C
NEPH RH 36.6 %
LED CURRENT 69.8 mA
REF DETECTOR 867 mV
RANGE AVG DIAGS ALARM
The CRn screen (read only) displays the current CRn reading.
●
In the Main Menu, choose Diagnostics >
CRn
.
Detector Status
Nephelometer
Status
CRn