Operation
Service Menu
3-98
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Service >
Nephelometer Calibration
.
NEPHELOMETER CALIBRATION:
>NEPH RH 4.0 %
NEPH TEMP 2.6
o
C
NEPH SRC LEVEL 35 %
RANGE AVG DIAGS ALARM
The Nephelometer RH screen allows the user to view and calibrate the
nephelometer relative humidity.
●
In the Main Menu, choose Service > Neph Calibration >
Neph RH
.
CALIBRATE NEPH RH:
NEPH RH: 50.0 %
OFFSET: 3.5 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Temperature screen allows the user to view and
calibrate the nephelometer temperature sensor calibration.
●
In the Main Menu, choose Service > Neph Calibration >
Neph
Temp
.
CALIBRATE NEPH TEMP:
NEPH TEMP: 25.0
o
C
OFFSET: 2.2
o
C
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Source Level screen allows the user to view and
calibrate the source level.
●
In the Main Menu, choose Service > Neph Calibration >
Neph Src
Level
.
Nephelometer RH
Nephelometer
Temperature
Nephelometer Source
Level