Operation
Service Menu
3-94
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
MASS COEFFICIENT:
>AUTO
MANUAL
RANGE AVG DIAGS ALARM
The Auto Mass Coefficient screen allows the user to conduct a mass foil
calibration.
Note
After pressing
a series of screens will walk you through the
mass foil calibration procedure. For details, see “
Auto Mass Coefficient
” on
page
4-9
.
▲
●
In the Main Menu, choose Service > Mass Calibration > Mass >
Auto
.
MASS COEF: 7000.0
FOIL VALUE: 144
2
3
μ
g
TO SAVE FOIL VALUE
AND OPEN BENCH
RANGE AVG DIAGS ALARM
The Manual Mass Coefficient screen allows the user to adjust the mass
coefficient used in beta attenuation.
●
In the Main Menu, choose Service > Mass Calibration > Mass >
Manual
.
MASS COEFFICIENT:
CURRENTLY: 7000
SET TO: 699
2
?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Alpha Efficiency Coefficient screen allows the user to view and set the
alpha efficiency value.
Auto Mass Coefficient
Manual Mass Coefficient
Alpha Efficiency
Coefficient