Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-87
CALIBRATE AMBIENT RH:
REL HUMIDITY: 40.0 %
OFFSET: 0.0 %
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure/Vacuum Calibration submenu allows the user to view and
calibrate the pressure sensor and vacuum flow sensor. The pressure/vacuum
calibration submenu is visible only when the instrument is in service mode.
For more information on the service mode, see “
Service Mode
” earlier in
this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Pres/Vacuum Calibration
.
PRES/VACUUM CALIBRATION:
>BARO PRES
VAC/FLOW
RANGE AVG DIAGS ALARM
The Barometer Pressure Calibration submenu is used to calibrate the
barometer pressure offset, span or reset default values.
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres
.
BARO PRES CALIBRATION:
>OFFSET 0.0
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
Pressure/Vacuum
Calibration
Barometer Pressure
Calibration