Operation
Volumetric Conditions
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-45
Note
The digital outputs may take up to one second after the assigned
state occurs to show up on the outputs.
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration
>
Output Relay Settings
.
OUTPUT RELAY SETTINGS:
>1 NOP CONC ALARM
2 NOP LOCAL/REMOTE
3 NOP UNITS
4 NOP GEN ALARM
5 NOP NONE
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital output pin and ground
is normally open and closes to trigger the digital output action.
●
Press
to toggle and set the logic state to normally open or
normally closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
The Instrument State submenu allows the user to select the instrument
state that is assigned to the selected relay output. A submenu lists signal
types of either alarms or non-alarm to choose from.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Output Relay Settings > select Relay >
Instrument State
.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM
Logic State
Instrument State
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......