C-Link Protocol Commands
Measurement Configuration
B-50
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Selection
g/m
3
mg/m
3
6 C1
C1
7 C2
C2
8 C3
C3
rh th
This command reports the current heater RH threshold for the sample.
The following example reports that the RH threshold is 30%.
Send:
rh
th
Receive:
rh th 30 %
set rh th
value
This command sets the heater RH threshold for the sample to
value
, where
value
is a floating-point number representation of RH threshold between
25 to 90%. The following example sets the RH threshold to 35%.
Send:
set rh th 35
Receive:
set rh th 35 ok
temp comp
This command reports whether temperature compensation is for actual or
standard conditions. The following example shows the temperature
compensation is standard.
Send:
temp
comp
Receive: temp
comp
std
set temp comp
selection
selection
= | act | std |
These commands turn the temperature compensation to either actual or
standard conditions. The following example turns the temperature
compensation to actual.
Send:
set temp comp act
Receive:
set temp comp act ok
temp std
This command reports the standard temperature. The following example
reports that the standard temperature is 18 °C.
Send:
temp
std
Receive:
temp std 18 degC
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......