Installation
Acceptance Testing and Startup Procedures
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
2-7
Figure 2–2.
Model 5030
i
Sampling Room Air Under Stable Conditions
Record the temperature sensor readings from the Model 5030
i
and
compare to your NIST-traceable thermometer. Each of these measurements
should be within ±2 °C tolerance of your NIST-traceable thermometer.
Within this tolerance, the temperature sensors have passed the acceptance
test.
If the sensors are slightly out of tolerance (±3 °C), the acceptance test
should be classified as marginal. If the sensor performance is outside this
range, please perform a temperature sensor calibration.
For more information about the temperature sensor calibration, see the
“
Calibration
” chapter.
Record the ambient RH sensor and nephelometer RH sensor readings from
the Model 5030
i
and compare to your NIST-traceable Hygrometer. The
Model 5030
i
RH sensors should compare within ±2% RH tolerance of
your NIST-traceable hygrometer. If the RH sensor performance is within
this tolerance, the acceptance test has passed.
Note
The ambient temperature verification should be completed prior to
performing the RH-sensor verification due to a thermal compensation
applied to the RH-sensor. Furthermore, try to avoid RH comparisons
<30% RH and >80% RH.
▲
One-Point Temperature
Verification
One-Point RH Sensor
Verification
Ta
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......