Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-101
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
●
In the Main Menu, choose Service > Analog Input Cal > select
Channel >
Calibrate Zero
.
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.04 V
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
●
In the Main Menu, choose Service > Analog Input Cal > select
Channel >
Calibrate Full Scale
. (Hook up a voltage source of 10 V
to the analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 9.84 V
SET TO: 10.0
0
V
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
Analog Input Calibrate
Zero
Analog Input Calibrate
Full-Scale
Содержание 5030i SHARP
Страница 4: ......
Страница 10: ...About This Manual Where to Get Help vi Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Страница 168: ......
Страница 218: ......
Страница 294: ......
Страница 306: ......
Страница 402: ......
Страница 403: ......