Servicing
Removing the SHARP Optics Assembly
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-11
The SHARP optics assembly can be removed from the measurement case
assembly. Refer to the following steps when a procedure requires lowering
the partition panel (
Figure 7–4
).
Figure 7–4.
Removing the SHARP Optics Assembly
Equipment Required:
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF and unplug the power cord.
2.
If the instrument is mounted in a rack, remove it from the rack.
3.
Disconnect the electrical connector.
4.
Unfasten the two captive hardware screws.
Removing the SHARP
Optics Assembly
Captive Screws (2)
Electrical Connector